    Stomatos vol.20 no.39 Canoas jul./dic. 2014

     

     

    Analysis of apical deviation by type of system used to drive NRT® MANI rotary instruments

     

    Análise do desvio apical de acordo com o tipo de sistema utilizado para acionamento dos instrumentos rotatórios NRT® MANI

     

     

    Elise Sasso Faccin I; Luiza Bonezi Boff II; Renata Grazziotin Soares III; Gustavo Golgo Kunert IV; Tiago André Fontoura de Melo V

     

    I Specialist in Endodontics from Faculdade São Leopoldo Mandic – Unidade Porto Alegre, Porto Alegre, RS, Brazil
    II Graduate student at Faculdade da Serra Gaúcha (FSG), Caxias do Sul, RS, Brazil
    III PhD in Endodontics from Universidade Federal do Rio Grande do Sul (UFRGS), Porto Alegre, RS, Brazil
    IV MSc in Endodontics from Faculdade São Leopoldo Mandic – Unidade Porto Alegre, Porto Alegre, RS, Brazil
    V MSc in Endodontics from Faculdade São Leopoldo Mandic – Unidade Porto Alegre, Porto Alegre, RS, Brazil, and a PhD in Endodontics from FSG, Caxias do Sul, RS, Brazil

    Correspondence

     

    The authors have no conflicts of interest to declare concerning the publication of this manuscript.

     

     


     

    ABSTRACT

    Purpose: One of the major challenges encountered in treatment of curved canals is maintaining their original features. This study aimed to examine whether the type of system used with NRT® rotary instruments might infl uence the occurrence of deviation in the apical region. Methods: Twenty simulated curved canals were divided into two experimental groups. In Group 1, canal preparation was performed with the pneumatic hand piece, using constant speed, but without torque control. In Group 2, an electric motor with constant speed and torque control was used. The simulated blocks were photographed before and after preparation using a platform to maintain the same position in both photographs. Images were manipulated in Adobe Photoshop® to evaluate deviation at 1 mm and 3 mm from the working length (WL). The t test was used for statistical analysis of data. The significance level was set at 1%. Results and Conclusion: Group 2 (electric motor) had lower values for deviation at both locations analyzed (1 mm and 3 mm from the WL). In both groups, sites 3 mm from the WL had lower mean deviation than sites 1 mm from WL.

    Keywords: Endodontics, Root Canal Preparation, Dental Equipment, Complications.


     

    RESUMO

    Objetivo: Um dos grandes desafios no tratamento de canais curvos é manter suas características originais. Este estudo teve como objetivo examinar se o tipo de sistema utilizado com os instrumentos rotatórios NRT® poderia infl uenciar a ocorrência de desvio na região apical. Métodos: Vinte canais curvos simulados foram divididos em dois grupos experimentais. No Grupo 1, a preparação dos canais foi realizada com um sistema pneumático, com velocidade constante mas sem controle de torque. No Grupo 2, utilizou-se um motor elétrico com velocidade constante e controle de torque. Os blocos simulados foram fotografados antes e após a preparação utilizando uma plataforma para que a mesma posição fosse mantida nas duas fotografias. As imagens foram manipuladas no software Adobe Photoshop® para avaliar o desvio a 1 mm e a 3 mm do comprimento de trabalho. O teste t foi usado para fazer a análise estatística dos dados. O nível de significância foi estabelecido em 1%. Resultados e conclusão: O Grupo 2 (motor elétrico) mostrou valores menores de desvio nos dois pontos analisados (1 mm e 3 mm do comprimento de trabalho). Em ambos os grupos, as áreas localizadas a 3 mm do comprimento de trabalho apresentaram desvio médio menor do que as áreas localizadas a 1 mm do comprimento de trabalho.

    Palavras-chave: Endodontia, Preparação de Canais Radiculares, Equipamentos Dentários, Complicações.


     

     

    INTRODUCTION

    The challenging when performing chemo-mechanical preparation of canals is to conduct endodontic instrumentation to enlarge the root canal that maintains its shape without causing deformation or deviation.

    According to Schäfer et al. 1 anatomical changes (deviations during root canals instrumentation) imposed on curved root canals during endodontic instrumentation may infl uence the prognosis of endodontic treatment success. Apical deviation compromises cleaning and elimination of microorganisms from the canal system, and this is particularly true of the apical third.

    With the advent of rotary nickel-titanium instruments (NiTi), occurrence of these accidents has been reduced. The reduction in such accidents is due to the properties of the alloy, such as fl exibility, torsional resistance and shape memory 2,3.

    Studies by Thompson and Dummer 4, Chen and Messer 5 and Schäfer et al. 6 showed that canal preparations performed with rotary NiTi instruments were more centralized in curved canals, in comparison with preparation performed with manual instruments. Additionally, the same authors demonstrated that it took less time to finish complete preparation when Ni-Ti instruments were used.

    However, some studies have shown that even using NiTi instruments it is not guaranteed that there will be no deviations during root canal preparation, given that other factors such as the experience of the operator 7, and the speed and the torque to which the electric motor used in activation is programmed can also have an influence.

    Rotary instruments are used at low speed and are driven by electric or pneumatic systems. Electric motors, commonly available on the market, provide accurate and constant control of speed and torque. However, Bortnik et al. 8 and Buchanan 9 have shown that there is no difference between electric and the compressed-air drive in terms of deformation or fracture of endodontic instruments.

    Therefore, the objective of this study was to examine whether the type of system used with NRT® rotary instruments might have an infl uence on the occurrence of deviation in the apical third of simulated canals.

     

    MATERIALS AND METHODS

    Twenty simulated canal blocks were used (Dentsply Maillefer, Ballaigues, VD, Switzerland), with 35º curvature, 16 mm length from the orifice and apical diameter of 0.15 mm.

    Canals were prepared with NRT® system rotary endodontic instruments (MANI Inc., Toshigi, Japan) in both experimental groups and the only variation in the preparation protocol was the choice of one of the two motors analyzed&rlquo; in the study (Table 1).

     

     

     

    During the canal preparation, simulated blocks were fixed to a counter mini clamp to facilitate instrumentation, with the direction of root canal curvature standardized as toward the operator's right hand. The working length (WL) was standardized at 15 mm.

    Canal preparation was performed by a single operator in both experimental groups and each set of instruments were used to prepare five simulated canals.

    Prior to and during preparation of the two experimental groups, canals were irrigated with distilled water at every change of endodontic instrument, (Iodontosul, Industrial Odontológica do Sul Ltda., Porto Alegre, Brazil) to remove resin debris, followed by anionic detergent irrigation with Tergensol (Inodon, Porto Alegre, Brazil) to lubricate the canal. Both irrigants were stored in a 10 mL disposable syringe (Plastipak Indústria Cirúrgica Ltda., Curitiba, Paraná, Brazil), fitted to a 25 x 04 hypodermic needle (Becton-Dickinson Indústria Cirúrgica Ltda., Curitiba, Paraná, Brazil). Aspiration was performed with a tip fitted to a size 40-20 cannula (Ibrás CBO Indústria Cirúrgica e Óptica S.A., Campinas, São Paulo, Brazil) which was in turn fitted to the suction unit of the dental chair (Gnatus Equipamentos Médico- Odontológicos Ltda., São Paulo, São Paulo, Brazil).

    The protocol for preparation of the simulated canals was the same in both experimental groups. Initially, cervical preparation was performed with stainless steel instruments, first #35/12 was taken to 5 mm, followed by #35/10 up to 7 mm from the entrance of the canal. Next, #25/06 and #30/06 were used at 3 mm from WL and then instruments #25/04 and #30/04 were used calibrated to the WL.

    In G1, the preparation of simulated canals was done with NRT® rotary instruments driven by an Endo Duo 40: 1 pneumatic piece (Adiel®, Ribeirão Preto, São Paulo, Brazil) coupled to a handpiece (Dabi Atlante, Ribeirão Preto, São Paulo, Brazil), with constant rotation of 250 rpm and no torque control.

    In the G2 group, canals were prepared with rotary NRT® instruments driven by an X-Smart electrical motor (Dentsply / Maillefer, Ballaigues, Switzerland), with constant rotation of 300 rpm and torque of 1.5 N and 3 N for instruments with tapers of 0.04 and 0.06, respectively. Instruments for cervical preparation with 0.10 and 0.12 tapers were used at a torque of 5 N, according to the manufacturer's recommendations.

    At every instrument change, a size #15 endodontic instrument (Dentsply/ Maillefer, Ballaigues, Switzerland) was manually inserted up to WL to remove resin debris from the apical area.

    Apical deviation was analyzed before and after preparation by positioning canal blocks on a platform and photographing them with a digital camera, always using the same object-to-focus distance. Indian ink was injected into root canals to improve the photographic contrast and aid viewing.

    Next, the images acquired were manipulated on Adobe Photoshop®, version 6.0. Images were transformed into millimeters by relating the original length of the canal to the image length on the computer screen, thereby avoiding reduction of the image pixels, maintaining the image's sharpness.

    The same software was used to adjust the contrast of the images. Each postoperative image was transformed into a layer with 50% transparency and then each one was superimposed onto the preoperative image. We were therefore able to observed both images superimposed as a result of the transparency.

    Next, the ruler tool was employed to locate the exact sites on the image at which deviations would be measured. These analysis points were pre-determined at 1 mm short of WL (Figure 1) and 3 mm short of WL (Figure 2).

     

     

     

     

     

    Deviations were measured at both points to be analyzed with the aid of the ruler tool. Distances were measured from the internal face (IF) of the original canal to the IF of the prepared canal and from the external face (EF) of the original canal to the EF of the prepared canal. Data were subjected to statistical analysis.

     

    RESULTS

    According to the t test, the group in which canals were prepared using an electric motor exhibited, on average, lower deviation at both analysis sites (3 mm and 1 mm of WL), when compared to canals instrumented with the pneumatic piece, to a 1% significance level (Table 2).

     

     

     

    The t test for paired observations showed that the site 3 mm from the WL had significantly lower mean deviation than the site at 1 mm from WL, to a 1% significance level (Figure 3).

     

     

     

    DISCUSSION

    During canal preparation one of the major challenges that dentists face is to maintain the original anatomy of the curved root canal and also the position of the apical foramen. Based on this challenge, the present study investigated whether the type of motor used to drive the NRT® system affects apical deviation in simulated canals.

    The use of ten samples in each group is in line with other studies, such as Thompson and Dummer 4 and Melo et al. 10. These authors also used simulated canals. One advantage of using simulated canals made from resin blocks is the possibility of standardizing the curvature, diameter and other features of the canals, such as, for example, the length of the straight part of the canal. This standardization reduces the possibility of bias in the results.

    For the present study, canals with 35 degrees of curvature were chosen, in common with a study by Schäfer et al. 6. Other studies have also used canals with abrupt curves, varying from 20 to 40 degrees 11-14. It is important and useful to exaggerate the clinical situation in order to assess how an endodontic instrument performs during preparation of severely curved canals.

    Comparison of the apical deviation in the two experimental groups showed that instruments driven by a pneumatic hand piece caused greater deviation than instruments driven by an electric motor. This finding could be explained by the fact that the electric motor allows an exact torque and speed to be chosen. In contrast, the pneumatic hand piece does not offer this choice of torque and speed, which may have directly infl uenced the higher values for apical deviation in Group 1.

    It is well known that many factors related to the properties of endodontic rotary instruments may act as adjuvants to apical deviation, such as, for example, the design of the instrument, its metallurgic composition, and variations in its tip diameter and taper 15. Based on this, our study is able to show that the rotary system is safer than the pneumatic system in terms of apical deviation, because of the ability to control the speed of rotation and torque, as reported by Limongi et al. 16 and López et al. 17.

    However, our findings are in disagreement with other autors, who showed that apical transportation and the ability to center the instrumented canal were not infl uenced by the type of instruments or type of movement: rotary or oscilatory 18,19. Capar et al. 20 also support this point of view, because they found that six different systems (including rotary and reciprocating) were similar in terms of the ability to center the instrumented canal and the likelihood of apical deviation. Additionally, Saber and Sadat 21 observed reduced apical transportation and better centered canals in molars that had been instrumented using WaveOne®.

    According to Dietz et al. 22, Yared et al. 23, Zelada et al. 24 and Gao et al. 25, when instruments are operated at higher activation speeds, there is increased risk of accidents during chemo-mechanical preparation of root canals and of damage to the rotary instrument, such as deformities and fractures.

    Safe use of devices (hand pieces) without speed controls to instrument canals is dependent on the operator's experience and mastery of the technique. According to Al-Omari et al. 26, an expert operator is able to prepare root canals with Rotary Ni-Ti instruments more quickly and more safely than non-expert dentists. Mandel et al. 27 highlight the need of previous mastery and training by the operator in relation to the chosen instruments and technique if canals are to be instrumented sucessfully. In-depth knowledge of how instruments should be used allied to previous experience may reduce endodontic acidentes and anatomic canal aberrations.

    The higher degree of apical deviation at 1 mm from the WL is in accordance with a study by Hata et al. 28. In this regard, Melo et al. 10 used the oscilating system and found higher mean deviation at 1 mm from the WL, compared with the deviation in the medium third of the canal (in the middle of the curvature). The literature has continued to show that the apical third, especially at 1 mm from the root end (or from the WL), is where the majority of apical deviation occurs.

    Therefore, it is very important to have in-depth knowledge about the instruments and the motor device chosen to instrument the canal. If it is not possible to control torque and speed during instrumentation of curved canals, then it will be very difficult to maintain the original shape of the canal.

     

    CONCLUSIONS

    According to the results of this study, it can be observed that:

    - Canal preparation with a pneumatic motor resulted in greater apical deviation when compared with canals instrumented using the electric motor.

    - The site at 1 mm from the WL exhibited greater apical deviation than the site at 3 mm from the WL.

     

    REFERENCES

    1. Schäfer E, Tepel J, Hoppe W. Properties of endodontic hand instruments used in rotary motion. Part 2. Instrumentation of curved canals. J Endod. 1995;21:493-7.         [ Links ]

    2. Walia HM, Brantley WA, Gerstein H. An initial investigation of the bending and torsional properties of nitinol root canal files. J Endod. 1988;14:346-51.

    3. Schäfer E. Root canal instruments for manual use: a review. Endod Dent Traumatol. 1997;13:51-64.

    4. Thompson SA, Dummer PM. Shaping ability of profile 04 taper series 20 rotatory nickel-titanium instruments in simulated root canals: part 1. Int Endod J. 1997;30:1-7.

    5. Chen JE, Messer HH. A comparison of stainless steel hand and rotary nickel-titanium instrumentation. Aust Dent J. 2002;47:12-20.

    6. Schäfer E, Erler M, Dammaschke T. Comparative study on the shaping ability and cleaning efficiency of rotary MTwo instruments: part 2. Cleaning effectiveness and shaping ability in severely curved root canals of extracted teeth. Int Endod J. 2006;39:203-12.

    7. Baumann MA, Roth A. Effect of experience on quality of canal preparation with rotary nickel-titanium files. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1999;88:714-8.

    8. Bortnik KL, Steiman HR, Ruskin A. Comparison of nickel-titanium file distortion using electric and air-driven handpiece. J Endod. 2001;27:57-9.

    9. Buchanan LS. The standardized-taper root canal preparation – part 2. File selection and safe handpiece – driven file use. Int Endod J. 2001;34:63-71. 10. Melo TAF, Weber A, Menon D, Soares RG, Salles AA. Análise da infl uência do grau de curvatura na ocorrência de desvios apicais após o prepare oscilatório em canais simulados. Rev Sul-Bras Odontol. 2010;7:312-9.

    11. Thompson SA, Dummer PM. Shaping ability of Hero 642 rotary nickel-titanium instruments in simulated root canals: Part 2. Int Endod J. 2000;33:255-61.

    12. Calberson FL, Deroose CA, Hommez GM, De Moor RJ. Shaping ability of ProTaper nickel-titanium files in simulated resin root canals. Int Endod J. 2004;37:613-23.

    13. Troian CH, Só MV, Figueiredo JA, Oliveira EPM. Deformation and fracture of RaCe and K3 endodontic instruments according to the number of uses. Int Endod J. 2006;39:616-25.

    14. Silva KT, Soares RG, Limongi O, Irala LED, Salles AA. Wear analysis promoted in simulated canals apical third after endodontic preparation with ProTaper Universal Sistem. J Appl Oral Sci. 2009;17:145-51.

    15. Kunert GG, Camargo Fontanella VR, Moura AA, Barletta FB. Analysis of apical root transportation associated with ProTaper Universal F3 and F4 instruments by using digital subtraction radiography. J Endod. 2010;36:1052-5.

    16. Limongi O, Klymus AO, Baratto Filho F, Vanni JR, Travassos R. In vitro evaluation of the presence of apical deviation with employment of automated handpieces with continuous and alternate motion for root canal preparation. J Appl Oral Sci. 2004;12:195-9.

    17. López FU, Fachin EV, Camargo Fontanella VR, Barletta FB, Só MV, Grecca FS. Apical transportation: a comparative evaluation of three root canal instrumentation techniques with three different apical diameters. J Endod. 2008;34:1545-8.

    18. Oliveira CA, Meurer MI, Pascoalato C, Silva SR. Cone-beam computed tomography analysis of the apical third of curved roots after mechanical preparation with different automated systems. Braz Dent J. 2009;20:376-81.

    19. Hartmann MS, Fontanella VR, Vanni JR, Fornari VJ, Barletta FB. CT evaluation of apical canal transportation associated with stainless steel hand files, oscillatory technique and pro taper rotary system. Braz Dent J. 2011;22:288-93.

    20. Capar ID, Ertas H, Ok E, Arslan H, Ertas ET. Comparative study of different novel nickel-titanium rotary systems for root canal preparation in severely curved root canals. J Endod. 2014;40:852-6.

    21. Saber Sel D, Abu El Sadat SM. Effect of altering the reciprocation range on the fatigue life and the shaping ability of WaveOne nickel-titanium instruments. J Endod. 2013;39:685-8.

    22. Dietz DB, Di Fiore MP, Bahcall JK, Lautenschlager EP. Effect of rotational speed on the breakage of nickel-titanium rotary files. J Endod. 2000;26:68-71.

    23. Yared GM, Bou Dagher FE, Machtou P. Infl uence of rotational speed, torque and operator's proficiency on ProFile failures. Int Endod J. 2001;34:47-53.

    24. Zelada G, Varela P, Martín B, Bahillo JG, Magán F, Ahn S. The effect of rotational speed and the curvature of root canals on the breakage of rotary endodontic instruments. J Endod. 2002;28:540-2.

    25. Gao Y, Shotton V, Wilkinson K, Phillips G, Johnson WB. Effects of raw material and rotational speed on the cyclic fatigue of ProFile Vortex rotary instruments. J Endod. 2010;36:1205-9.

    26. Al-Omari MA, Aurich T, Wirtti S. Shaping canals with ProFiles and K3 instruments: does operator experience matter? Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2010;110:e50-5.

    27. Mandel E, Adib-Yazdi M, Benhamou LM, Lachkar T, Mesgouez C, Sobel M. Rotary Ni-Ti profile systems for preparing curved canals in resin blocks: infl uence of operator on instrument breakage. Int Endod J. 1999;32:436-43.

    28. Hata G, Uemura M, Kato AS, Imura N, Novo NF, Toda T. A comparison of shaping ability using ProFile, GT File and Flex-R endodontic instruments in simulated canals. J Endod. 2002;28:316-21.

     